Добавить новость
News in English


Новости сегодня

Новости от TheMoneytizer

Repeatability, reproducibility, agreement, and safety of Tono-Pen tip cover for intraocular measurement using latex and polyethylene wrap

by Pukkapol Suvannachart, Somkiat Asawaphureekorn, Sunee Chansangpetch, Abhibol Inobhas, Krit Pongpirul

Purpose

To evaluate repeatability, reproducibility, and agreement of intraocular pressure measurement with Tono-Pen using Ocufilm and polyethylene wrap tip cover in human eyes.

Methods

This is a cross-sectional, experimental study. A gas-sterilized, polyethylene wrap was used as an alternative for Tono-Pen tip cover. For the right eye, 4 measurements using polyethylene wrap tip cover were done by two examiners (A and B) in random order to assess intra-observer repeatability and inter-observer reproducibility. For the left eye, 4 measurements were done by examiner A using both polyethylene wrap tip cover and Ocufilm in random order to assess intra-observer repeatability and agreement. Bland-Altman plot and intraclass correlation coefficient (ICC) were used in all analyses. Cost minimization analysis was evaluated.

Results

For examiner A, the repeatability of polyethylene wrap tip cover was -0.34, 95% limits of agreement (LOA) were -3.04 to 2.36, and ICC was 0.93 in the right eyes. As for the left eyes, the repeatability of polyethylene wrap tip cover was -0.33, 95% LOA were -3.01 to 2.36, and ICC was 0.93. For examiner B, the repeatability of polyethylene wrap tip cover was -0.02, 95% LOA were -2.88 to 2.83, and ICC was 0.92. The inter-observer reproducibility of polyethylene wrap tip cover was 0.36, 95% LOA were -3.34 to 4.07, and ICC was 0.90. The repeatability of Ocufilm was -0.42, 95% LOA were -2.75 to 1.91, and ICC was 0.95. The agreement of polyethylene wrap tip cover and Ocufilm was -0.71, 95% LOA were -5.18 to 3.76, and ICC was 0.83. There were no allergic reactions or serious complications. From the cost minimization analysis, the local cost for polyethylene tip cover was approximately 8 times lower compared to Ocufilm.

Conclusions

Tono-pen with Ocufilm and polyethylene wrap tip cover were used to measure the intraocular pressure. The polyethylene wrap tip cover demonstrated acceptable repeatability, reproducibility, and agreement with Ocufilm in normotensive eyes, and had a good safety profile.

Читайте на сайте


Smi24.net — ежеминутные новости с ежедневным архивом. Только у нас — все главные новости дня без политической цензуры. Абсолютно все точки зрения, трезвая аналитика, цивилизованные споры и обсуждения без взаимных обвинений и оскорблений. Помните, что не у всех точка зрения совпадает с Вашей. Уважайте мнение других, даже если Вы отстаиваете свой взгляд и свою позицию. Мы не навязываем Вам своё видение, мы даём Вам срез событий дня без цензуры и без купюр. Новости, какие они есть —онлайн с поминутным архивом по всем городам и регионам России, Украины, Белоруссии и Абхазии. Smi24.net — живые новости в живом эфире! Быстрый поиск от Smi24.net — это не только возможность первым узнать, но и преимущество сообщить срочные новости мгновенно на любом языке мира и быть услышанным тут же. В любую минуту Вы можете добавить свою новость - здесь.




Новости от наших партнёров в Вашем городе

Ria.city
Музыкальные новости
Новости России
Экология в России и мире
Спорт в России и мире
Moscow.media










Топ новостей на этот час

Rss.plus