Precision and consistency of the passive leg raising maneuver for determining fluid responsiveness with bioreactance non-invasive cardiac output monitoring in critically ill patients and healthy volunteers

by Sahil Chopra, Jordan Thompson, Shahab Shahangian, Suman Thapamagar, Dafne Moretta, Chris Gasho, Avi Cohen, H. Bryant Nguyen

Objective

The passive leg raising (PLR) maneuver has become standard practice in fluid resuscitation. We aim to investigate the precision and consistency of the PLR for determining fluid responsiveness in critically ill patients and healthy volunteers using bioreactance non-invasive cardiac output monitoring (NiCOM™, Cheetah Medical, Inc., Newton Center, Massachusetts, USA).

Methods

This study is prospective, single-center, observational cohort with repeated measures in critically ill patients admitted to the medical intensive care unit and healthy volunteers at a tertiary academic medical center. Three cycles of PLR were performed, each at 20–30 minutes apart. Fluid responsiveness was defined as a change in stroke volume index (ΔSVI) > 10% with each PLR as determined by NiCOM™. Precision was the variability in ΔSVI after the 3 PLR’s, and determined by range, average deviation and standard deviation. Consistency was the same fluid responsiveness determination of “Yes” (ΔSVI > 10%) or “No” (ΔSVI ≤ 10%) for all 3 PLR’s.

Results

Seventy-five patients and 25 volunteers were enrolled. In patients, the precision was range of 17.2±13.3%, average deviation 6.5±4.0% and standard deviation 9.0±5.2%; and for volunteers, 17.4±10.3%, 6.6±3.8% and 9.0±6.7%, respectively. There was no statistical difference in the precision measurements between patients and volunteers. Forty-nine (65.3%) patients vs. twenty-four (96.0%) volunteers had consistent results, p < 0.01. Among those with consistent results, twenty-four (49.0%) patients and 24 (100%) volunteers were fluid responsive.

Conclusions

The precision and consistency of determining ΔSVI with NiCOM™ after PLR may have clinical implication if ΔSVI > 10% is the absolute cutoff to determine fluid responsiveness.

Читайте на сайте


Smi24.net — ежеминутные новости с ежедневным архивом. Только у нас — все главные новости дня без политической цензуры. Абсолютно все точки зрения, трезвая аналитика, цивилизованные споры и обсуждения без взаимных обвинений и оскорблений. Помните, что не у всех точка зрения совпадает с Вашей. Уважайте мнение других, даже если Вы отстаиваете свой взгляд и свою позицию. Мы не навязываем Вам своё видение, мы даём Вам срез событий дня без цензуры и без купюр. Новости, какие они есть —онлайн с поминутным архивом по всем городам и регионам России, Украины, Белоруссии и Абхазии. Smi24.net — живые новости в живом эфире! Быстрый поиск от Smi24.net — это не только возможность первым узнать, но и преимущество сообщить срочные новости мгновенно на любом языке мира и быть услышанным тут же. В любую минуту Вы можете добавить свою новость - здесь.